The most crucial process to assess in a tumor for efficient management is neo-angiogenesis. Currently, no reliable biomarker is available to measure angiogenesis, and direct microvessel density assay requires frequent biopsies for serial tumor monitoring. Technological innovations in imaging, however, have enabled the capture of physiological information in addition to anatomic details. One promising imaging technique for acquiring physiological information about angiogenesis is perfusion CT, which provides anatomic information as well.
The most crucial process to assess in a tumor for efficient management is neo-angiogenesis. Currently, no reliable biomarker is available to measure angiogenesis, and direct microvessel density assay requires frequent biopsies for serial tumor monitoring. Technological innovations in imaging, however, have enabled the capture of physiological information in addition to anatomic details. One promising imaging technique for acquiring physiological information about angiogenesis is perfusion CT,1 which provides anatomic information as well.
Initial reports on dynamic perfusion CT date back to 1980,2 But the limitations of earlier CT scanners and a lack of availability of analytic software restricted this technique to use as a research tool. Recent developments in multislice CT and new availability of commercial perfusion software, however, have led to a resurgence of clinical interest in perfusion CT, with rapid expansion of its applications. The clinical role of perfusion CT in stroke management is well established, but its role in oncology for diagnosis, risk stratification, and therapeutic response monitoring continues to evolve.
PERFUSION CT FOR ONCOLOGY
Clinical management in oncology shifted after the concept of tumor neo-angiogenesis was introduced. This process of new blood vessel growth is critical for tumor growth and metastasis.3 Newer generations of oncologic treatment regimens target neo-angiogenesis, delivering less systemic toxicity than did conventional cytotoxic agents. Anti-angiogenic drugs are targeted specifically to the growth factors that stimulate neo-angiogenesis.
In tumor tissue, the release of several abnormal mediators such as vascular endothelial growth factor (VEGF)4 leads to development of a new network of capillaries. Angiogenic vasculature differs in several aspects from that in normal mature tissue. Blood vessels in tumors are typically dilated and tortuous, with abnormal branching patterns and dead ends, and they lack organized structure into arterioles, capillaries, and venules.3 These tumor blood vessels have increased permeability over normal tissue.3
Angiogenic factors increase microvessel density, which is defined as the mean number of endothelial cells or endothelial cell clusters in a given number of microscopic fields that are clearly separate from adjacent tumor cells and connective tissue elements.5 Microvessels are highlighted by staining the endothelial cells with anti-CD34 antibodies.
Microvessel density has been established as a prognostic indicator for many cancers, and the most direct strategy to monitor anti-angiogenic therapy would therefore be periodic biopsy. This approach is invasive, however, and likely unacceptable to patients. Moreover, estimates of microvessel density may not reflect the overall tumor biology if a single-site biopsy is used, due to the heterogeneity of perfusion within the tumor.
Noninvasive imaging, on the other hand, can be repeated in patients with larger volume coverage than biopsies. Imaging studies allow assessment of physiological parameters-such as blood volume, blood flow, permeability, and mean transit times-that reflect tumor biology. Imaging may therefore be superior to microvessel density measurement for assessing angiogenesis.6 Perfusion CT involves dynamic scanning after administration of iodinated contrast material followed by mathematic modeling to study contrast material kinetics in the tissue. Estimates of tumor perfusion parameters such as blood flow, blood volume, mean transit time, and permeability-surface area product can be derived by applying kinetic models to the CT perfusion data set (see table).
Among the various mathematic models that have evolved for performing perfusion studies, the deconvolution method is currently the most validated and widely used approach. Perfusion CT involves acquisition of unenhanced images followed by a series of postcontrast dynamic images in axial cine mode with static table position. The unenhanced images are obtained to select the region of interest for dynamic imaging of tumor perfusion. Because current CT techniques (four- to 64-slice CT) allow scanning of only 2 to 4 cm of tumor tissue, careful selection of region of interest is key. Breath-hold is crucial for perfusion CT study of upper abdominal viscera where breathing artifacts are more likely. Duration of scanning ranges from 30 seconds to two minutes from arrival of contrast, depending on clinical need, organ of interest, and analytical method.
Time of arrival of contrast may vary depending on the location of the tumor. Contrast would arrive earlier in a lung mass, for example, than a mass in the foot. Earlier reports recommended scanning for at least two minutes when permeability measurement was required,7 but current recommendations advocate shorter durations. A recent study reported that a 65-second scan is adequate for measurement of permeability of colorectal carcinoma.8 Radiation dose, breath-hold demands, and patient motion invariably limit the maximum duration of scanning. There is always a trade-off between radiation dose and temporal resolution.
ONCOLOGIC APPLICATIONS
Clinical imaging tools for evaluation of angiogenesis should help the oncologist select optimum treatment by tailoring the dose of anti-angiogenic drugs, detecting early anti-angiogenic response, and analyzing impact of treatment on tumor progression.
Data on body tumor imaging with perfusion CT are emerging in studies of various applications. One drawback is the wide variability of scanning techniques and mathematical modeling used in these studies. Furthermore, most of the studies use a small sample. Readers should be aware of these limitations before drawing conclusions from these studies and implementing these applications in routine practice.
It has been shown that in healthy individuals, CT-derived hepatic arterial and portal venous perfusion values are inversely correlated, whereas in a setting of liver metastases, there is a direct positive correlation between these values.11 Miles et al reported that increased peritumoral arterial perfusion and increased global liver perfusion indicate a more favorable outcome following chemotherapy.9 Tsushima et al reported that HCC has nine times greater hepatic arterial perfusion than background liver parenchyma, whereas colorectal carcinoma metastases demonstrate only four times greater arterial perfusion.12 A few patients with HCC showed a decrease in hepatic perfusion values following transarterial chemoembolization.
Miles et al reported a decrease in pancreatic perfusion in patients with type 1 diabetes mellitus and failing pancreatic transplant, whereas pancreatic perfusion increased in insulinoma and Wilson's disease compared with healthy individuals.13 Another study reported increased blood flow and blood volume in insulinoma compared with background pancreatic parenchyma. Time to peak enhancement and tumor permeability were comparable to normal pancreas, however.14 Even in the absence of disease in the pancreas, it has been reported that pancreatic perfusion decreases with advancing age.15 Abe et al validated pancreatic perfusion CT by showing excellent agreement of pancreatic tumor blood flow measurement between perfusion CT and xenon CT.16
It is important to emphasize that none of the perfusion CT acquisition or analytical techniques are standardized for implementation in routine clinical practice for oncologic body imaging, and all of the perfusion parameters calculated by commercial software are only estimates and not absolute values. One of the principal limitations of CT perfusion is the limited scanning volume. Current four- to 16-slice CT scanners allow 2-cm tissue coverage for dynamic CT acquisition. Even with a 64-slice scanner, the maximum coverage is only 4 cm. Hence, only a portion of the tumor or organ can be sampled for perfusion measurements. Scan volume selection is thus a crucial aspect of scanning technique.
Another issue is patient motion, which may be corrected by registration within the image plane. Patient motion out of image plane, however, causes data loss and errors in perfusion values. Respiratory gating may improve motion problems at the expense of temporal resolution.25 Adequate patient instruction for breath-hold, with immobilization straps over the abdomen, may help.
Beam-hardening artifacts from metallic stents, prostheses, and surgical implants can also result in variation in the perfusion values, and careful selection of region of interest for scanning is important here as well. Likewise, adequate distention of hollow viscus such as colorectum or stomach with saline or any neutral contrast is crucial for optimum perfusion measurements. In our institution, we use 250 to 300 mL of saline to distend the rectum before conducting perfusion CT studies.
Radiation dose is an important concern in implementing perfusion CT in routine clinical practice. Currently, the radiation dose is two to four times greater for perfusion CT than routine contrast-enhanced CT of the same region using the same scanning parameters, depending on the scanner.
FUTURE DIRECTIONS
Efforts to improvise further perfusion CT applications are ongoing. Respiratory gating, for example, would reduce misregistration breathing artifacts. Measurement of peak tissue enhancement can provide more physiological data by normalizing them for the patient's body weight and dose of contrast medium administered. Such an approach could determine the standardized perfusion value (SPV), defined as the ratio of tumor perfusion to mean whole-body perfusion.26 Mean whole-body perfusion is cardiac output divided by patient weight.
Miles et al reported that SPV measurements correlated with standardized uptake value measured on FDG-PET for lung nodules. The threshold of SPV to distinguish between benign and malignant lesions would be 1.5.26 The increasing availability of PET/CT raises the possibility of combining CT perfusion imaging with PET as well. This combined PET/CT/perfusion "triple" imaging could obtain anatomic, metabolic, and angiogenic information about tumor with a single scan.
Some vendors have recently introduced a table toggle technique to increase scan coverage and address current limitations in scanning volume in perfusion CT. In this technique, the scanner couch toggles between two locations for dynamic acquisition. This application is only possible, however, at the expense of temporal resolution, which would be five seconds instead of one second in static table technique.
Currently available iodinated intravenous contrast materials are limited by their short intravascular life. Further research should develop new agents with a longer intravascular life that could overcome these difficulties. Using automatic exposure control and low tube potential could optimize radiation dose during serial dynamic scanning.
Perfusion CT offers new opportunities for detection, risk stratification, and therapeutic monitoring of tumors. Although extensively researched in current literature, the clinical role of perfusion CT in the body is still evolving. Further research and improvements in acquisition and analytic techniques would open doors for perfusion CT to become a routine protocol for oncologic imaging and expand its clinical applications.
Dr. Namasivayam is a research fellow, and Dr. Sahani is director of CT, both at Massachusetts General Hospital in Boston. Dr. Sahani has received grants and research support from GE Healthcare.
References1. Miles KA, Charnsangavej C, Lee FT, et al. Application of CT in the investigation of angiogenesis in oncology. Acad Radiol 2000;7:840-850.
2. Axel L. Cerebral blood flow determination by rapid-sequence computed tomography: theoretical analysis. Radiology 1980;137:679-686.
3. Folkman J. New perspectives in clinical oncology from angiogenesis research. Eur J Cancer 1996; 2A:2534-2539.
4. Fontanini G, Vignati S, Boldrini L, et al. Vascular endothelial growth factor is associated with neovascularization and influences progression of non-small cell lung carcinoma. Clin Cancer Res 1997;3:861-865.
5. Tae K, El-Naggar AK, Yoo E, et al. Expression of vascular endothelial growth factor and microvessel density in head and neck tumorigenesis. Clin Cancer Res 2000;6:2821-2828.
6. Marx J. Angiogenesis. A boost for tumor starvation. Science 2003; 301:452-454.
7. Miles K, Dawson P, Blomley M. Functional computed tomography. Oxford, UK: Isis Medical Media, 1997.
8. Goh V, Halligan S, Hugill JA, et al. Quantitative colorectal cancer perfusion measurement using dynamic contrast-enhanced multidetector-row computed tomography: effect of acquisition time and implications for protocols. JCAT 2005;29:59-63.
9. Dugdale PE, Miles KA. Hepatic metastases: the value of quantitative assessment of contrast enhancement on computed tomography. Eur J Radiol 1999;30:206-213.
10. Fukushima H, Itoh S, Takada A, et al. Diagnostic value of curved multiplanar reformatted images in multislice CT for the detection of resectable pancreatic ductal adenocarcinoma. Eur Radiol 2006;16:1709-1718.
11. Miles KA, Leggett DA, Kelley BB, Hayball MP, Sinnatamby R, Bunce I. In vivo assessment of neovascularization of liver metastases using perfusion CT. Br J Radiol 1998;71:276-281.
12. Tsushima Y, Funabasama S, Aoki J, et al Quantitative perfusion map of malignant liver tumors, created from dynamic computed tomography data. Acad Radiol 2004;11:215-223.
13. Miles KA, Hayball MP, Dixon AK. Measurement of human pancreatic perfusion using dynamic computed tomography with perfusion imaging. Br J Radiol 1995;68:471-475.
14. Xue HD, Jin ZY, Liu W, et al. [Perfusion characteristics of normal pancreas and insulinoma on multi-slice spiral CT]. Zhongguo Yi Xue Ke Xue Yuan Xue Bao 2006;28:68-70.
15. Tsushima Y, Kusano S. Age-dependent decline in parenchymal perfusion in the normal human pancreas: measurement by dynamic computed tomography. Pancreas 1998;17:148-152.
16. Abe H, Murakami T, Kubota M, et al. Quantitative tissue blood flow evaluation of pancreatic tumor: comparison between xenon CT technique and perfusion CT technique based on deconvolution analysis. Radiat Med 2005;23:364-370.
17. Goh V, Halligan S, Gartner L, et al. Quantitative colorectal cancer perfusion measurement by multidetector-row CT: does greater tumour coverage improve meaurement reproducibility? Br J Radiol 2006;79:578-583.
18. Sahani DV, Kalva SP, Hamberg LM, et al. Assessing tumor perfusion and treatment response in rectal cancer with multisection CT: initial observations. Radiology 2005;234:785-792.
19. Willett CG, Boucher Y, di Tomaso E, et al. Direct evidence that the VEGF-specific antibody bevacizumab has antivascular effects in human rectal cancer. Nat Med 2004;10:145-147.
20. Willett CG, Boucher Y, Duda DG, et al. Surrogate markers for antiangiogenic therapy and dose-limiting toxicities for bevacizumab with radiation and chemotherapy: continued experience of a phase I trial in rectal cancer patients. J Clin Oncol 2005;23:8136-8139.
21. Henderson E, Milosevic MF, Haider MA, Yeung IW. Functional CT imaging of prostate cancer. Phys Med Biol 2003;48:3085-3100.
22. Harvey CJ, Blomley MJ, Dawson P, et al. Functional CT imaging of the acute hyperemic response to radiation therapy of the prostate gland: early experience. JCAT 2001;25:43-49.
23. Haider MA, Milosevic M, Fyles A, et al. Assessment of the tumor microenvironment in cervix cancer using dynamic contrast enhanced CT, interstitial fluid pressure and oxygen measurements. Int J Radiat Oncol Biol Phys 2005;62:1100-1107.
24. Dugdale PE, Miles KA, Bunce I, et al. CT measurement of perfusion and permeability within lymphoma masses and its ability to assess grade, activity, and chemotherapeutic response. JCAT 1999;23:540-547.
25. Miles KA, Griffiths MR. Perfusion CT: a worthwhile enhancement? Br J Radiol 2003;76:220-231.
26. Miles KA, Griffiths MR, Fuentes MA. Standardized perfusion value: universal CT contrast enhancement scale that correlates with FDG PET in lung nodules. Radiology 2001;220:548-553.
Study Reaffirms Low Risk for csPCa with Biopsy Omission After Negative Prostate MRI
December 19th 2024In a new study involving nearly 600 biopsy-naïve men, researchers found that only 4 percent of those with negative prostate MRI had clinically significant prostate cancer after three years of active monitoring.
Study Examines Impact of Deep Learning on Fast MRI Protocols for Knee Pain
December 17th 2024Ten-minute and five-minute knee MRI exams with compressed sequences facilitated by deep learning offered nearly equivalent sensitivity and specificity as an 18-minute conventional MRI knee exam, according to research presented recently at the RSNA conference.
Can Radiomics Bolster Low-Dose CT Prognostic Assessment for High-Risk Lung Adenocarcinoma?
December 16th 2024A CT-based radiomic model offered over 10 percent higher specificity and positive predictive value for high-risk lung adenocarcinoma in comparison to a radiographic model, according to external validation testing in a recent study.